Repairs
We provide quick repair turnaround time and warranty our workmanship.
Nationwide Supply
Nationwide supply of durable, new and professionally refurbished equipment.
Servicing & Maintenance
We are proficient with medical electronics PCB Repairs, servicing & Maintenance.
AMC Contract
we are undertaking the repair and maintenance job most all the medical equipment.
The Vivo 40 is the ideal pressure ventilator for transition from hospital to home, and for home ventilation thanks to an integrated re-chargeable battery which gives upto 4 hours of backup and a Target volume feature.This, again, is FDA approved for invasive ventilation and for non-invasive respiratory support with the choice of PSV, PCV, or CPAP modes and a pediatric mode.
With the Target volume feature the delivered inspiratory pressure will automatically be adjusted to optimize the delivered tidal volume to the patient, adjust to the patients lung mechanics and reduce the need for ventilator adjustments.
Specifications:
• Patient Modes:Adult, Paediatric (>10 kg)
• Device Modes: Clinical, Home
• IPAP/EPAP: 4 to 40 cmH2O / 2 to 20 cmH2O or IPAP -2 cmH2O
• CPAP: 4 to 20 cmH2O
• Breath Rate: 4 to 40 BPM
• Inspiratory Time: 0.3 to 5 s
• Rise Time: 1 to 9
• Inspiratory Trigger (eSync): 1 to 9, Off
• Expiratory Trigger: 1 to 9
• Target Volume: 200-1500 ml
• Ramp function setting: 10 to 60 min, Off (PCV, PSV and CPAP mode)
• Humidification setting: 1 to 9 (10 to 30 mgH2O/L)
• Audible alarm setting: 1 to 9
• Maximum flow:> 200 litres/min
• Minimal Inspiratory Time Off: 0.3 to 3.0 s
• Maximal Inspiratory Time: 0.3 to 3.0 s, Off. (0.3 to in Paediatric Mode)
• Noise level (at 10 cmH2O constant pressure): Less than 30 dB(A)
• Indications: Pressure; bar graph and digital data: 0 to 40 cmH2O
• Calculated tidal volume accuracy: ± 20% or ± 20 ml (whichever is the greatest)
• Leakage calculation accuracy: ± 20%
• Total Breath Rate: 0 to 60 BPM (Breaths Per Minute)
• I/E Ratio: 1 :10 to 10 :1
• Alarms
Audible Alarm Level: 35 to 85 dB(A)
• Low pressure alarm, high pressure alarm, low volume alarm, low breath rate alarm, high breath rate alarm, low leakage alarm, high leakage alarm, low internal battery warning, low power alarm, internal function failure alarms.
Power Supplies
• Mains supply: 100 to 240 V AC
• Internal battery / External battery: Capacity 3.8 Ah / 12/24 V DC
Operating Conditions
• Recommended mask leakage: 20 to 40 litres/min at 10 cmH2O
• Minimum mask leakage: 12 litres/min at 4 cmH2O
Dimensions
• W x H x D: 190 x 243 x 223 mm (without HA20)
• Weight: 4.0 kg (without HA20)