The Vivo 30 is an advanced and clinically safe respiratory solution, which lends itself to usage as a home ventilator. The Vivo 30 provides non-invasive respiratory support through Pressure Control and Pressure Support modes. Thanks to the proprietary technology eSync – the Vivo 30 synchronizes smoothly with the patient's breathing pattern.
The Minimal Inspiratory Time in PSV mode the Vivo 30 prevents premature termination of inspiratory pressure. Conversely, the Maximal Inspiratory Time in PSV mode the Vivo 30 prevents prolonged inspiratory pressure, to customize therapy and to prevent late cycling off.Its large screen and intuitive user-interface with home and clinical modes make it one of the most user-friendly home ventilators available.
Specifications:
• Patient Modes: Adult, Paediatric (>10 kg)
• Device Modes: Clinical, Home
• IPAP/EPAP: 4 to 40 cmH2O / 2 to 20 cmH2O or IPAP -2 cmH2O
• CPAP: 4 to 20 cmH2O
• Breath Rate: 4 to 40 BPM
• Inspiratory Time: 0.3 to 5 s
• Rise Time: 1 to 9
• Inspiratory Trigger (eSync): 1 to 9, Off
• Expiratory Trigger: 1 to 9
• Target Volume: 200-1500 ml
• Ramp function setting: 10 to 60 min, Off (PCV, PSV and CPAP mode)
• Humidification setting: 1 to 9 (10 to 30 mgH2O/L)
• Audible alarm setting: 1 to 9
• Maximum flow:> 200 litres/min
• Minimal Inspiratory Time Off: 0.3 to 3.0 s
• Maximal Inspiratory Time: 0.3 to 3.0 s, Off. (0.3 to in Paediatric Mode)
• Noise level (at 10 cmH2O constant pressure): Less than 30 dB(A)
• Indications: Pressure; bar graph and digital data: 0 to 40 cmH2O
• Calculated tidal volume accuracy: ± 20% or ± 20 ml (whichever is the greatest)
• Leakage calculation accuracy: ± 20%
• Total Breath Rate: 0 to 60 BPM (Breaths Per Minute)
• I/E Ratio: 1 :10 to 10 :1
• Alarms
Audible Alarm Level: 35 to 85 dB(A)
• Low pressure alarm, high pressure alarm, low volume alarm, low breath rate alarm, high breath rate alarm, low leakage alarm, high leakage alarm, low internal battery warning, low power alarm, internal function failure alarms.
Power Supplies
• Mains supply: 100 to 240 V AC
• Internal battery / External battery: Capacity 3.8 Ah / 12/24 V DC
Operating Conditions
• Recommended mask leakage: 20 to 40 litres/min at 10 cmH2O
• Minimum mask leakage: 12 litres/min at 4 cmH2O
Dimensions
• W x H x D: 190 x 243 x 223 mm (without HA20)
• Weight: 4.0 kg (without HA20)
• Air outlet: 22 mm male conical standard connector